Model No. |
Capacity |
Liters |
Usable Space |
ALS-BOD-4 | 4cu.ft. | 113 | 610X450X410mm |
ALS-BOD-6 | 6cu.ft. | 173 | 845X500X400mm |
ALS-BOD-10 | 10cu.ft. | 282 | 870X570X560mm |
ALS-BOD-12 | 12cu.ft. | 342 | 900X650X570mm |
ALS-BOD-15 | 15cu.ft. | 421 | 700X900X650mm |
ADDITIONAL FACILITIES AT AN EXTRA COST of BOD incubators: – |
Lighting system by interior illumination with 3 fluorescent tubes of 60cm. (20 Watts placed Vertically along back wall.) |
Timer with reverses 0 – 24 hours for regulating hours of light and darkness including fitting at the top of incubator |
Voltage Stabilizer |
Technical specification of BOD Incubators PID Based |
|||
HSN Code:90272000 | Temperature Range: 10℃ to 60℃ | Accuracy: 0.2℃ | Uniformity:2.0℃ |
CONTROLLER SYSTEM | Make : Anachem Lab microprocessor based Pid controller with dry-rtd type pt-100 sensor with printer interface- storage printing or online | ||
BODY & DOOR INSULATION | Poly Urethane Foam (PUF) 75 MM THICK | ||
CHAMBER ILLUMINATION | Led Lighting is provided in the Chamber | ||
HEATING SYSTEM | S.S. Nichrome Wire Embedded ‘U’ Shaped Air Heater. | ||
REFRIGERATION SYSTEM | Cooling System Consists Of Condenser & Compressor. | ||
COMPRESSOR | Hermetically Sealed Compressor | ||
CONTROLLER | Delta Electronics Make Temp. Controller | ||
SENSOR | RTD TYPE PT-100 Temp. Sensor (CLASS-A) | ||
SAMPLE SECURITY | Fermod, France Make Door Latch with Key & Hinges | ||
GLASS DOOR | Glass door For View The Inside |
Safety Cut Off: – A High Temperature Thermostat is provided for as an additional safety against Temperature overshoot & cut off the heater supply |
Principle of BOD Incubator
A Biological Oxygen Demand (BOD) Incubator is a crucial tool in Environmental and wastewater management. It helps determine the level of organic pollution in water by Measuring the amount of oxygen consumed by Microorganisms during the degradation of organic matter. The incubator operates based on a few fundamental principles.
Controlled Temperature
BOD Incubators maintain a constant and precise temperature, usually set at 20°C, to mimic environmental conditions. This controlled environment ensures consistency in BOD testing.
Aerobic Microbial Activity
BOD incubators create an aerobic atmosphere by providing a continuous supply of oxygen to support microbial activity. This promotes the breakdown of organic substances in water samples.
Monitoring Dissolved Oxygen
A dissolved oxygen sensor constantly measures the oxygen level, Allowing for the Calculation of BOD values over a Specified Incubation period, typically five days.
Accurate Assessment
By following these principles, BOD Incubators Facilitate Accurate and Standardized Assessment of water quality, aiding in Environmental protection and Regulatory Compliance.
Reviews